Calibrate in Air 20.9%
Hold sensor with plug | | | |
|
 | Part number | Step2 | Product name
| Assembly |
Assemble DIN Kit.
Oxygen reading should still read 20.9% +/- 1 on the last digit. | | | |
|
 | Part number | Step3 | Product name
| Open cylinder |
Gently open cylinder until a soft hiss is heard.
Leave open for 5 seconds then turn the cylinder off. | | | |
|
 | Part number | Step4 | Product name
| Wait |
After 10 seconds turn cylinder on again for 5 second.
The reading may go higher, then drop back, due to flow generating pressure. | | | |
|
 | Part number | Step5 | Product name
| Take Reading |
When reading stops dropping the O2 level has been measured.
IF in doubt repeat from step 3. | | | |
|
 | Part number | Step6 | Product name
| Finished |
Always check calibration has not moved.
If it has moved repet from step 1. | | | |
|